Spectrometer Features:
- Microwave Frequency: CW X-Band (9.0 - 9.8 GHz).
- X-band Solid State Ultra Low Noise Microwave Bridge (200 mW max leveled source output, 60 dB max attenuation, 60 MHz tuning range).
- Integrated Bridge Controller with real time display (automated tuning and cavity matching). Full Ethernet control of all functions.
- Digital Ultra High Resolution Hall Field Controller (100 G – 18 kG operating range, up to 0 – 18 kG in one scan sweep range).
- High Linearity Modulation Amplifier Module: 500 Hz – 120 kHz
- High Sensitivity Probehead with optical window for CW-EPR in X-band (up to 10 mm sample tube access, 20 G max modulation amplitude, automated cavity matching, unloaded Q > 15,000, sensitivity weak pitch: 1800 : 1).
- Liquid helium cryostat: 4 - 100 K.
|
Experiments:
- Room Temperature
- Time Scans (kinetics studies)
- 2D (a second parameter can be varied along with the magnetic field)
- Cryogenic experiments:
- + VT unit, using liquid nitrogen, range: 100 – 500 K.
- + Liquid helium cryostat, range: 4 – 100 K.
|